This article was posted 02/19/2007 and is most likely outdated.

Important Safety Notice and Recall of the Amprobe
 

 
Topic - Safety
Subject
- Important Safety Notice and Recall of the Amprobe

February 19, 2007  

Ask a Question |  Weekly Code GraphicQuizzes |  Free Stuff InstructorsOnline Training Products | Seminars | SubscribeUnsubscribe
[ image1 Post Comments | View Comments | Notify Me When Comments Are Added ] Web Page Version [Printer-Friendly]    

Important Safety Notice and Recall of the Amprobe
ACD-10 PRO, ACD-10 TRMS-PRO, ACD-14 and ACD-14 TRMS Digital Clamp Meters

January 15, 2007
Dear Amprobe Customer:

In cooperation with the U.S. Consumer Product Safety Commission, Amprobe is voluntarily recalling the ACD-10 PRO, ACD-10 TRMS-PRO, ACD-14 and ACD-14 TRMS Digital Clamp Meters. Please stop using all of these products and return your unit to Amprobe and your unit will be replaced with an AC75B or ACD-15 TRMS PRO Digital Clamp Meter

Description of the Problem
The ACD-10 PRO, ACD-10 TRMS-PRO, ACD-14 and ACD-14 TRMS models "lock up" in certain circumstances failing to properly measure hazardous voltage levels on electrical systems. When the clamp meter locks up it provides readings under 1 volt or a millivolt reading. If the user does not follow industry standard practice by verifying the tester before and after checking a circuit, he/she may believe there is no voltage present when in fact there is. THIS COULD PLACE THE USER IN A POTENTIALLY HAZARDOUS SITUATION.

Requested Actions
Stop using your ACD-10 PRO, ACD-10 TRMS-PRO, ACD-14 or ACD-14 TRMS Clamp Meter and contact Amprobe directly for a free replacement clamp meter. A postage-paid return mailer and a return tracking number can be requested through the worldwide web (use the Continue button below) -OR- by calling 1-800-350-8661 Monday through Friday between 7 a.m. and 4 p.m. Pacific Time.

As a replacement clamp, we offer either AC75B or ACD-15 TRMS-PRO. The features of the replacement clamps are provided in the table below.

Image

Please return your affected ACD-10 and/or ACD-14 Digital Clamp Meter(s) as soon as possible, even if you have not experienced problems. If you are not the primary user of an ACD-10 or ACD-14, please pass this notice along to the appropriate people within your organization.

We recognize that recalls inconvenience our customers and we are committed to providing you with an alternate product with minimal disruption for you. If you have any additional questions about this recall, please call 1-800-350-8661.

Sincerely,
Jonathan Morrow
Amprobe Business Unit Director

 

[ View More Newsletters ] [ Send to a Friend ] [ Post Comments | View Comments | Notify Me When Comments Are Added ]

Copyright © Mike Holt Enterprises, Inc. All rights reserved.
This article is protected by United States copyright and other intellectual property laws and may not be
displayed or published on the internet without the prior written permission of Mike Holt Enterprises, Inc.

http://www.MikeHolt.com     1-888-NEC-CODE (1-888-632-2633)

Experiencing a Problem? Contact our Webmaster

Comments
    No comments to display

Get notified when new comments are posted here
* Your Email:
 
        
 
Add Your Comments to this Newsletter
* Your Name:
   Your name will appear under your comments.

* Your Email:
   Your email address is not displayed.
* Comments:

Email Notification Options:
Notify me when a reply is posted to this comment
Notify me whenever a comment is posted to this newsletter